452User Manual 1337.9952.02 ─ 12
For serial protocol options, search on a serial bus is available. These searches have
protocol-specific search criteria. For details, see the relevant chapters of the "Protocol
Analysis" chapter.
Edge......................................................................................................................452
Glitch.....................................................................................................................453
Width.....................................................................................................................453
Runt.......................................................................................................................454
Window................................................................................................................. 455
Timeout................................................................................................................. 456
Interval.................................................................................................................. 457
Slew rate............................................................................................................... 458
Setup & Hold.........................................................................................................459
State......................................................................................................................461
Pattern...................................................................................................................461
11.2.2.1 Edge
The edge search works the same way as the edge trigger.
Slope
Sets the edge type: rising edge ("Positive"), falling edge ("Negative"), or both.
Remote command:
SEARch:TRIGger:EDGE:SLOPe on page 1485
Level
Sets the voltage level for trigger search. The value is used for all search event types
that require one trigger level. The search trigger level is search-specific, you can define
different levels in different searches for the same event.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1484