Replacing the Pipette Assembly
7-6 DSX™ System Service Manual
3 Remove the Pipette Assembly from the Z-Drive Assembly by removing the two
M3 x 12 SHCS screws that attach the Pipette Assembly to the Drive Assembly
(Figure 7-3). The block is pinned in the center, so rocking the pipette from left to
right will frequently aid in removing the module from the Z mount.
Figure 7-3 Mounted Pipette Assembly