The Electronics Pod Assembly (EPOD)
17-10 DSX™ System Service Manual
17.3.3.4. Removing and Replacing the XY Printed Circuit Board
The XY printed circuit board (Part No. 528050004) is used in the movement of the X
motor and the Y motor (either motor can be driven individually or both can be driven
together). The board is mounted on top of the communications board as shown in
Figure 17-8.
Figure 17-8 XY Printed Circuit Board
To remove the XY Printed Circuit Board, remove the four M3 X 6mm cap head screws
and four M3 washers