GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-10 Section 8-3 - System Software - Installation/Upgrade Procedure
8-3-5-2-1-5 Connectivity - Recording Device Destination settings
1.) From the touch panel, select Utility -> Connectivity -> Service.
Figure 8-3 Connectivity - Recording Device Destination settings
2.) Record all settings in Table 8-7 for each Destination Device and Service.
8-3-5-2-1-6 Connectivity - Recording Dataflow settings
1.) From the touch panel, select Utility -> Connectivity -> Dataflow.
Table 8-7 Record settings from Service screen
Destination Device Properties
Table 8-6 Record settings for each device listed in Device table
Name IP Address