K117sm7e3
7-44
11. Tap [Quit] to close Test screen.
12. Press [4 Points Calibration].
13. On Calibration screen, a blinking X symbol on the bottom left can be seen.
Press the X until it stops blinking with a beep.
NOTE
Press the X symbol for several seconds before the progress bar at the bottom reaches the
end.