Remote control commands
R&S
®
RTP
1208User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<PrbAttMdAuto> Range: 1E-3 to 1000
*RST: 1
Default unit: V/V
Usage: Query only
Asynchronous command
PROBe<m>:SETup:OFFSet:AZERo
Performs an automatic correction of the zero error. If the DUT is ground-referenced,
the AutoZero function can improve the measurement results.
See also: "Detect AutoZero, Use AutoZero" on page 166
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
PROBe<m>:SETup:OFFSet:USEautozero <AutoZeroOffset>
Includes the AutoZero offset in measurement results. The auto zero error is detected
with PROBe<m>:SETup:OFFSet:AZERo.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<AutoZeroOffset> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: Version 2.70
CHANnel<m>:EATScale <ExtAttScl>
Sets the attenuation scale for an external divider.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAttScl> LIN | LOG
*RST: LIN