Remote control commands
R&S
®
RTP
2110User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:SWIRe:DATA:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:SWIRe:DATA:MIN and
TRIGger<m>:SWIRe:DATA:MAX.
*RST:
EQUal
TRIGger<m>:SWIRe:DATA:MAX
<DataPatternMax>
Sets the the end value of a data for the data pattern range if TRIGger<m>:SWIRe:
DATA:CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<DataPatternMax>
TRIGger<m>:SWIRe:DATA:MIN <DataPatternMin>
Specifies a data for the data pattern, or sets the the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<DataPatternMin>
TRIGger<m>:SWIRe:ERRor:ESC <ErrorESC>
Enables searching for an escape error.
Suffix:
<m>
.
1..3
Parameters:
<ErrorESC> ON | OFF
*RST: ON