Remote control commands
R&S
®
RTP
2111User Manual 1337.9952.02 ─ 12
TRIGger<m>:SWIRe:ERRor:PARity <ErrorParity>
Enables triggering on a parity error (even parity).
Suffix:
<m>
.
1..3
Parameters:
<ErrorParity> ON | OFF
*RST: ON
TRIGger<m>:SWIRe:TIME:CONDition <TimeOperator>
Set the condition for the data value for the time code. You can define an exact value or
a value range.
Suffix:
<m>
.
1..3
Parameters:
<TimeOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:SWIRe:TIME:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:SWIRe:TIME:MIN and
TRIGger<m>:SWIRe:TIME:MAX.
*RST:
EQUal
TRIGger<m>:SWIRe:TIME:MAX <TimePatternMax>
Sets the the end value of a data value for the time code range if TRIGger<m>:
SWIRe:TIME:CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<TimePatternMax>
TRIGger<m>:SWIRe:TIME:MIN <TimePatternMin>
Specifies a pattern for the data value for the time code, or sets the the start value of a
pattern range.
Suffix:
<m>
.
1..3
Protocols