Mixed signal option (MSO, R&S RTP-B1)
R&S
®
RTP
945User Manual 1337.9952.02 ─ 12
Remote command:
TRIGger<m>:PARallel:SETHold:HTIMe on page 2213
14.3.1.6 State
The state trigger detects the logical state of several logically combined digital channels
at a given clock edge. The trigger source is a logical combination of digital channels or
a parallel bus. The trigger occurs at the clock edge at which the state condition is true.
Figure 14-5: State trigger settings for trigger source = parallel bus
Clock source
Selects the digital channel of the clock signal.
Remote command:
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue] on page 2210
TRIGger<m>:PARallel:STATe:CSOurce:VALue on page 2210
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue] on page 2210
Clock edge
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time at which the logical states and the bus value are analyzed.
Remote command:
TRIGger<m>:PARallel:STATe:CSOurce:EDGE on page 2214
Channel states
Available for "Trigger Source"= "Par. Bus1/2/3/4"
For each digital channel that is used in the bus, set the required state: 1, 0, or X (don't
care).
Trigger