Remote control commands
R&S
®
RTP
2213User Manual 1337.9952.02 ─ 12
TRIGger<m>:PARallel:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Usage: Asynchronous command
22.18.4.5 Setup & Hold trigger
See also:
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue] on page 2210
TRIGger<m>:PARallel:SETHold:CSOurce:EDGE............................................................ 2213
TRIGger<m>:PARallel:SETHold:HTIMe..........................................................................2213
TRIGger<m>:PARallel:SETHold:STIMe..........................................................................2213
TRIGger<m>:PARallel:SETHold:CSOurce:EDGE <ClockEdge>
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time reference point for the setup and hold time measurement.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:PARallel:SETHold:HTIMe <HoldTime>
Sets the minimum time after the clock edge while data should be stable and not
change its state.
See also: "Hold time" on page 944
Parameters:
<HoldTime> Range: -99.8E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:PARallel:SETHold:STIMe <SetupTime>
Sets the minimum time before the clock edge while data should be stable and not
change its state.
Mixed signal option (MSO, R&S
RTP-B1)