Remote control commands
R&S
®
RTP
1798User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:TBTO:DATA:DCONdition....................................................................1800
SEARch:TRIGger:TBTO:DATA:DMAX............................................................................1800
SEARch:TRIGger:TBTO:DATA:DMIN.............................................................................1801
SEARch:TRIGger:TBTO:DATA:ICONdition..................................................................... 1801
SEARch:TRIGger:TBTO:DATA:IMAX............................................................................. 1801
SEARch:TRIGger:TBTO:DATA:IMIN.............................................................................. 1802
SEARch:TRIGger:TBTO:ERRor:CRC.............................................................................1802
SEARch:TRIGger:TBTO:ERRor:FEC............................................................................. 1802
SEARch:TRIGger:TBTO:ERRor:OOR............................................................................ 1802
SEARch:TRIGger:TBTO:ERRor:ZERO...........................................................................1803
SEARch:TRIGger:TBTO:LENGth:CONDition.................................................................. 1803
SEARch:TRIGger:TBTO:LENGth:MAX...........................................................................1803
SEARch:TRIGger:TBTO:LENGth:MIN............................................................................1804
SEARch:TRIGger:TBTO:SADDress:CONDition...............................................................1804
SEARch:TRIGger:TBTO:SADDress:MAX....................................................................... 1804
SEARch:TRIGger:TBTO:SADDress:MIN........................................................................ 1805
SEARch:TRIGger:TBTO:TYPE......................................................................................1805
SEARch:TRIGger:TBTO:CRC:CONDition <SearchName>,<CRCOperator>
SEARch:TRIGger:TBTO:CRC:CONDition? <SearchName>
Sets the condition for the frame check. You can define an exact value or a value range.
Parameters:
<CRCOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:TRIGger:TBTO:
CRC:MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:TBTO:CRC:MIN and
SEARch:TRIGger:TBTO:CRC:MAX.
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:TBTO:CRC:MAX <SearchName>,<CRCPatternMax>
SEARch:TRIGger:TBTO:CRC:MAX? <SearchName>
Sets the end value of the frame check range if SEARch:TRIGger:TBTO:CRC:
CONDition is set to INRange or OORange.
Parameters:
<CRCPatternMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
Protocols