Spectrum analysis
R&S
®
RTP
417User Manual 1337.9952.02 ─ 12
Zoom ← Gate Definition
Zoom coupling is available if a zoom is defined. As long as "Zoom coupling" is
enabled, the gate area is defined identically to the zoom area - if you change the zoom,
the gate changes as well.
If several zoom diagrams are defined, select the zoom diagram to be used for gating.
The "Start" and "Stop" values of the gate are adjusted accordingly.
Zoom coupling can be set for measurement gates, FFT gates, and search gates. The
zoom must be defined on the diagram that contains the signal source of the measure-
ment, FFT, or search.
Remote command:
MEASurement<m>:GATE:ZCOupling on page 1425
MEASurement<m>:GATE:ZDIagram on page 1426
CALCulate:MATH<m>:FFT:GATE:ZCOupling on page 1451
SEARch:GATE:ZCOupling on page 1510
SEARch:GATE:ZDIagram on page 1510
Gate description ← Gate Definition
Defines whether the gate settings are configured using absolute or relative values.
"Absolute"
The gate is defined by absolute start and stop values.
"Relative"
The gate's start and stop values are defined by a percentage of the
value range.
Remote command:
CALCulate:MATH<m>:FFT:GATE:MODE on page 1450
MEASurement<m>:GATE:MODE on page 1424
SEARch:GATE:MODE on page 1508
(Relative) Start ← Gate Definition
Defines the starting value for the gate.
Remote command:
CALCulate:MATH<m>:FFT:GATE:ABSolute:STARt on page 1450
CALCulate:MATH<m>:FFT:GATE:RELative:STARt on page 1451
MEASurement<m>:GATE:ABSolute:STARt on page 1424
MEASurement<m>:GATE:RELative:STARt on page 1424
SEARch:GATE:ABSolute:STARt on page 1509
SEARch:GATE:RELative:STARt on page 1509
(Relative) Stop ← Gate Definition
Defines the end value for the gate.
Remote command:
CALCulate:MATH<m>:FFT:GATE:ABSolute:STOP on page 1450
CALCulate:MATH<m>:FFT:GATE:RELative:STOP on page 1451
MEASurement<m>:GATE:ABSolute:STOP on page 1424
MEASurement<m>:GATE:RELative:STOP on page 1424
SEARch:GATE:ABSolute:STOP on page 1509
SEARch:GATE:RELative:STOP on page 1510