Series 2600B System SourceMeter® instrument Reference Manual Section 4: Triggering
2600BS-901-01 Rev. F/August 2021 4-41
Controlling digital I/O lines
Although the digital I/O lines are primarily intended for use with a device handler for limit testing, they
can also be used for other purposes, such as controlling external logic circuits. You can control lines
either from the front panel or over a remote interface.
To set digital I/O values from the front panel:
1. Press the MENU key, select DIGOUT, and then press the ENTER key or press the
navigation wheel.
2. Select DIG-IO-OUTPUT, and then press the ENTER key or the navigation wheel.
3. Set the decimal value as required to set digital I/O lines in the range of 0 to 16,383 (see the table
in Digital I/O bit weighting (on page 4-41)), and then press the ENTER key or the
navigation wheel. For example, to set digital I/O lines 3 and 8, set the value to 132.
4. Press the EXIT (LOCAL) key as needed to return to the main menu.
To write-protect specific digital I/O lines to prevent their values from being changed:
1. Press the MENU key, then select DIGOUT and then press the ENTER key or the
navigation wheel.
2. Select WRITE-PROTECT, and then press the ENTER key or the navigation wheel.
3. Set the decimal value as required to write-protect digital I/O lines within the range of 0 to 16,383
(see Digital I/O bit weighting (on page 4-41)), and then press the ENTER key or the navigation
wheel. For example, to write-protect digital I/O lines 4 and 10, set the value to 520.
4. Press the EXIT (LOCAL) key as needed to return to the main menu.
To remove write protection, reset the decimal value to include only the lines that you want to write
protect. To remove write protection from all lines, set the value to 0.
Digital I/O bit weighting
Bit weighting for the digital I/O lines is shown in the following table.
Digital bit weight
Line #
Bit
Decimal weighting
Hexadecimal weighting
1
B1
1
0x0001
2
B2
2
0x0002
3
B3
4
0x0004
4
B4
8
0x0008
5
B5
16
0x0010
6
B6
32
0x0020
7
B7
64
0x0040
8
B8
128
0x0080
9
B9
256
0x0100
10
B10
512
0x0200
11
B11
1,024
0x0400
12
B12
2,048
0x0800
13
B13
4,096
0x1000
14
B14
8,192
0x2000

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals