Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-222 2600BS-901-01 Rev. F/August 2021
Also see
Measure auto delay (on page 2-45)
reset() (on page 9-177)
smuX.measure.count (on page 9-220)
smuX.measure.delayfactor (on page 9-222)
smuX.source.delay (on page 9-241)
smuX.reset() (on page 9-236)
smuX.measure.delayfactor
This attribute stores a multiplier to the delays that are used when smuX.measure.delay is set to
smuX.DELAY_AUTO.
Instrument reset
SMU reset
Recall setup
Usage
delayFactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayFactor
The delay factor multiplier
Source-measure unit (SMU) channel (for example, smua.measure.delayfactor
applies to SMU channel A)
Details
The delay factor is only applied when smuX.measure.delay = smuX.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic
delay.
Setting this attribute to zero disables delays when smuX.measure.delay = smuX.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
Measure auto delay (on page 2-45)
reset() (on page 9-177)
smuX.measure.delay (on page 9-221)
smuX.reset() (on page 9-236)