Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-392 2600BS-901-01 Rev. F/August 2021
print(tsplink.trigger[1].overrun)
If an event was ignored, displays true; if an event was not ignored, displays false.
Also see
tsplink.trigger[N].assert() (on page 9-388)
tsplink.trigger[N].clear() (on page 9-388)
tsplink.trigger[N].mode (on page 9-390)
tsplink.trigger[N].release() (on page 9-393)
tsplink.trigger[N].reset() (on page 9-393)
tsplink.trigger[N].stimulus (on page 9-394)
tsplink.trigger[N].wait() (on page 9-395)
tsplink.trigger[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers. This command is not
available on the 2604B, 2614B, or 2634B.
Instrument reset
TSP-Link trigger N reset
Recall setup
Usage
width = tsplink.trigger[N].pulsewidth
tsplink.trigger[N].pulsewidth = width
The pulse width (in seconds)
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20 μs.
Also see
tsplink.trigger[N].release() (on page 9-393)