Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-377
trigger.timer[N].delaylist
This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Not saved
10e-6 (10 µs)
Usage
intervals = trigger.timer[N].delaylist
trigger.timer[N].delaylist = intervals
intervals
Table of delay intervals in seconds
N
Trigger timer number (1 to 8)
Details
Each time the timer is triggered after it is enabled, it uses the next delay period from the array. The
default value is an array with one value of 10 µs.
After all elements in the array have been used, the delays restart at the beginning of the list.
If the array contains more than one element, the average of the delay intervals in the list must
be ≥ 50 µs.
Example
trigger.timer[3].delaylist = {50e-6, 100e-6, 150e-6}
DelayList = trigger.timer[3].delaylist
for x = 1, table.getn(DelayList) do
print(DelayList[x])
end
Set a delay list on trigger timer 3 with three delays (50 µs, 100 µs, and 150 µs).
Read the delay list on trigger timer 3.
Output:
5e-05
0.0001
0.00015
Also see
trigger.timer[N].reset() (on page 9-379)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals