EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #597 background image
Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-276 2600BS-901-01 Rev. F/August 2021
Also see
smuX.trigger.arm.set() (on page 9-254)
smuX.trigger.autoclear (on page 9-257)
smuX.trigger.endpulse.set() (on page 9-259)
smuX.trigger.initiate() (on page 9-263)
smuX.trigger.measure.set() (on page 9-265)
Triggering (on page 4-1)
smuX.trigger.source.stimulus
This attribute defines which event causes the source event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0
Usage
eventID = smuX.trigger.source.stimulus
smuX.trigger.source.stimulus = eventID
eventID
Set to the event that triggers the end-pulse source off action
X
Source-measure (SMU) channel (for example, smua.trigger.source.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event. When set, the
SMU waits for the event at the source event detector portion of the trigger model. To bypass waiting
for an event, set the value of this attribute to zero (0). Set eventID to one of the existing trigger
event IDs shown in the following table.
Trigger event IDs*
Event ID
Event description
smuX.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU) transitions
from the idle state to the arm layer of the trigger model
smuX.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to the
trigger layer of the trigger model
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement action
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smuX.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[N].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[N].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[N].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object N
display.trigger.EVENT_ID
Occurs when the TRIG key on the front panel is pressed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals