Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-223
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1
Usage
filterCount = smuX.measure.filter.count
smuX.measure.filter.count = filterCount
filterCount
The number of readings required for each filtered measurement (1 to 100)
X
Source-measure unit (SMU) channel (for example,
smua.measure.filter.count applies to SMU channel A)
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
smua.measure.filter.type = smua.FILTER_MOVING_AVG
smua.measure.filter.enable = smua.FILTER_ON
Sets the filter count for SMU channel A to 10.
Sets the filter type to moving average.
Enables the filter.
Also see
Filters (on page 2-66)
reset() (on page 9-177)
setup.recall() (on page 9-197)
smuX.measure.filter.enable (on page 9-224)
smuX.measure.filter.type (on page 9-225)
smuX.reset() (on page 9-236)