Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-144 2600BS-901-01 Rev. F/August 2021
lan.trigger[N].assert()
This function simulates the occurrence of the trigger and generates the corresponding event ID.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[N].assert()
N
The LAN event number (1 to 8)
Details
Generates and sends a LAN trigger packet for the LAN event number specified.
Sets the pseudo line state to the appropriate state.
The following indexes provide the listed LXI events:
â–ª 1:LAN0
â–ª 2:LAN1
â–ª 3:LAN2
▪ …
â–ª 8:LAN7
Example
lan.trigger[5].assert()
Creates a trigger with LAN packet 5.
Also see
lan.lxidomain (on page 9-134)
lan.trigger[N].mode (on page 9-148)
lan.trigger[N].overrun (on page 9-149)
lan.trigger[N].stimulus (on page 9-151)
lan.trigger[N].wait() (on page 9-152)
Understanding hardware value and pseudo line state (on page 4-23)
lan.trigger[N].clear()
This function clears the event detector for a LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[N].clear()
N
The LAN event number (1 to 8) to clear

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals