Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-135
Example
print(lan.lxidomain)
Displays the LXI domain.
Also see
None
lan.nagle
This attribute controls the state of the LAN Nagle algorithm.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
0 (lan.DISABLE)
Usage
state = lan.nagle
lan.nagle = state
state
The state of the Nagle algorithm:
â–ª 1 or lan.ENABLE: Enable the LAN Nagle algorithm for TCP connections
â–ª 0 or lan.DISABLE: Disable the Nagle algorithm for TCP connections
Details
This attribute enables or disables the use of the LAN Nagle algorithm on transmission control protocol
(TCP) connections.
Also see
lan.restoredefaults() (on page 9-136)
lan.reset()
This function resets the LAN interface.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.reset()
Details
This function resets the LAN interface. It performs the commands lan.restoredefaults() and
lan.applysettings(). It also resets the LAN password.
Also see
lan.applysettings() (on page 9-125)
lan.restoredefaults() (on page 9-136)
localnode.password (on page 9-156)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals