Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-162 2600BS-901-01 Rev. F/August 2021
Details
This indicates the instrument serial number.
Example
display.clear()
display.settext(localnode.serialno)
Clears the instrument display.
Places the serial number of the instrument on the top line of its display.
Also see
localnode.description (on page 9-154)
localnode.model (on page 9-156)
localnode.revision (on page 9-161)
localnode.showerrors
This attribute sets whether or not the instrument automatically sends generated errors.
Usage
errorMode = localnode.showerrors
localnode.showerrors = errorMode
Show error setting:
â–ª Show errors: 1
â–ª Do not show errors: 0
Details
If this attribute is set to 1, the instrument automatically sends any generated errors stored in the error
queue, and then clears the queue. Errors are processed after executing a command message (just
before issuing a prompt if prompts are enabled).
If this attribute is set to 0, errors are left in the error queue and must be explicitly read or cleared.
When using this command from a remote node, replace localnode with the node reference, for
example, node[5].showerrors.
Example
Enables sending of generated errors.
Also see
localnode.prompts (on page 9-158)