Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-255
smuX.trigger.arm.stimulus
This attribute selects the event that causes the arm event detector to enter the detected state.
SMU reset
Instrument reset
Recall setup
Usage
eventID = smuX.trigger.arm.stimulus
smuX.trigger.arm.stimulus = eventID
Event that triggers the arm detector
Source-measure unit (SMU) channel (for example, smua.trigger.arm.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event.
Set this attribute to zero to bypass waiting for events at the arm event detector (the SMU continues
uninterrupted through the remote trigger model). Set eventID to one of the existing trigger event IDs
shown in the following table.
smuX.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU) transitions from
the idle state to the arm layer of the trigger model
smuX.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to the trigger
layer of the trigger model
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement action
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smuX.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[N].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[N].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
Occurs when the appropriate LXI trigger packet is received on
LAN trigger object N
Occurs when the TRIG key on the front panel is pressed
Occurs when a *TRG command is received on the remote
interface
GPIB only: Occurs when a GET bus command is received
USB only: Occurs when a USBTMC TRIGGER message is
received
VXI-11 only: Occurs with the VXI-11 command
device_trigger; reference the VXI-11 standard for
additional details on the device trigger operation
trigger.blender[N].EVENT_ID
Occurs after a collection of events is detected
trigger.timer[N].EVENT_ID
Occurs when a delay expires
trigger.generator[N].EVENT_ID
Occurs when the trigger.generator[N].assert()
function is executed