Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-270 2600BS-901-01 Rev. F/August 2021
smuX.trigger.source.action
This attribute enables or disables sweeping the source (on or off).
SMU reset
Instrument reset
Recall setup
Usage
action = smuX.trigger.source.action
smuX.trigger.source.action = action
Sweep source action. Set to one of the following values:
â–ª 0 or smuX.DISABLE: Do not sweep the source
â–ª 1 or smuX.ENABLE: Sweep the source
Source-measure unit (SMU) channel (for example,
smua.trigger.source.action applies to SMU channel A)
Details
This attribute is used to enable or disable source level changes during a sweep. In addition to
enabling the action before initiating the sweep, make sure to configure it using
smuX.trigger.source.linearY(), smuX.trigger.source.listY(), or
smuX.trigger.source.logY().
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
smua.trigger.source.action = smua.ENABLE
Configure list sweep for SMU channel A (sweep through 3 V, 1 V, 4 V, 5 V, and 2 V).
Enable the source action.
Also see
smuX.trigger.source.linearY() (on page 9-272)
smuX.trigger.source.listY() (on page 9-273)
smuX.trigger.source.logY() (on page 9-274)
Triggering (on page 4-1)