Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-370 2600BS-901-01 Rev. F/August 2021
Details
Indicates if an event was ignored because the event detector was already in the detected state when
the event occurred. This is an indication of the state of the event detector that is built into the event
blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any
other trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output is true.
If an event was not ignored, the output is false.
Also see
trigger.blender[N].reset() (on page 9-370)
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Usage
trigger.blender[N].reset()
The trigger event blender (up to six)
Details
The trigger.blender[N].reset() function resets the following attributes to their factory
defaults:
• trigger.blender[N].orenable
• trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Example
trigger.blender[1].reset()
Resets the trigger blender 1 settings to factory defaults.
Also see
trigger.blender[N].orenable (on page 9-369)
trigger.blender[N].overrun (on page 9-369)