bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Usage
statusInformation = bufferVar.statuses[N]
The status value when reading N of the specified buffer was acquired
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in
the buffer
Details
This read-only buffer recall attribute is like an array (a Lua table) of the status values for all the
readings in the buffer. The status values are floating-point numbers that encode the status value; see
the following table for values.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.
Over temperature condition
Measure range was autoranged
Source range was autoranged
4-wire (remote) sense mode enabled
Relative offset applied to reading
Source function was limited because the complementary
function would be over the compliance limit