Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-407
tspnet.tsp.abort(testConnection)
Stops remote instrument execution on testConnection.
Also see
None
tspnet.tsp.abortonconnect
This attribute contains the setting for abort on connect to a TSP-enabled instrument.
Instrument reset
Recall setup
Usage
tspnet.tsp.abortonconnect = value
value = tspnet.tsp.abortonconnect
â–ª Enable: 1
â–ª Disable: 0
Details
This setting determines if the instrument sends an abort message when it attempts to connect to a
TSP-enabled instrument using the tspnet.connect() function.
When you send the abort command on an interface, it causes any other active interface on that
instrument to close. If you do not send an abort command (or if tspnet.tsp.abortonconnect is
set to 0) and another interface is active, connecting to a TSP-enabled remote instrument results in a
connection. However, the instrument does not respond to subsequent reads or executes because
control of the instrument is not obtained until an abort command has been sent.
Example
tspnet.tsp.abortonconnect = 0
Configure the instrument so that it does not send an abort command when connecting to a
TSP-enabled instrument.
Also see
tspnet.connect() (on page 9-399)