EasyManua.ls Logo

Tektronix KEITHLEY 2600B Series

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-220 2600BS-901-01 Rev. F/August 2021
Example
smua.measure.calibratev(1, 1e-4, 1e-5, 0.92, 0.903)
Adjust SMU channel A voltage measurement using the following values:
1 V calibration range
1e−4 for +zero measurement reading
1e−5 for +zero DMM measurement reading
0.92 for +FS measurement reading
0.903 for the +FS DMM measurement reading
Also see
Adjustment (on page 10-15)
smuX.cal.lock() (on page 9-204)
smuX.cal.restore() (on page 9-206)
smuX.cal.save() (on page 9-207)
smuX.cal.unlock() (on page 9-209)
smuX.source.calibrateY() (on page 9-239)
smuX.measure.count
This attribute sets the number of measurements made when a measurement is requested.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1
Usage
count = smuX.measure.count
smuX.measure.count = count
count
Number of measurements
X
Source-measure unit (SMU) channel (for example, smua.measure.count applies
to SMU channel A)
Details
This attribute controls the number of measurements made any time a measurement is requested.
When using a reading buffer with a measure command, this attribute also controls the number of
readings to be stored.
If the count is set to a value greater than 1, any measurement delay set by smuX.measure.delay
only occurs before the first measurement, while the smuX.measure.interval controls the interval
between successive measurements.
Example
smua.measure.count = 10
Sets the SMU channel A measure count to 10.

Table of Contents

Related product manuals