Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-371
trigger.blender[N].stimulus[M]
This attribute specifies the events that trigger the blender.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Trigger blender N reset
Not applicable
trigger.EVENT_NONE
Usage
eventID = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = eventID
eventID
The event that triggers the blender action; see Details
N
An integer representing the trigger event blender (up to six)
M
An integer representing the stimulus index (1 to 4)
Details
There are four stimulus inputs that can each select a different event. The eventID parameter can be
the event ID of any trigger event.
Use zero to disable the blender input.
The eventID parameter may be one of the existing trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
Event description
smuX.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU) transitions
from the idle state to the arm layer of the trigger model
smuX.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to the
trigger layer of the trigger model
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement action
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smuX.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[N].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[N].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[N].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object N
display.trigger.EVENT_ID
Occurs when the TRIG key on the front panel is pressed
trigger.EVENT_ID
Occurs when a *TRG command is received on the
remote interface
GPIB only: Occurs when a GET bus command is
received
USB only: Occurs when a USBTMC TRIGGER message
is received
VXI-11 only: Occurs with the VXI-11 command
device_trigger; reference the VXI-11 standard for
additional details on the device trigger operation
trigger.blender[N].EVENT_ID
Occurs after a collection of events is detected
trigger.timer[N].EVENT_ID
Occurs when a delay expires

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals