Series 2600B System SourceMeter® instrument Reference Manual Section 8: Instrument programming
2600BS-901-01 Rev. F/August 2021 8-97
2. Send the command:
connectionID = tspnet.connect(ipAddress)
Where:
â–ª connectionID is the connection ID that is used as a handle in all other TSP-Net function
calls.
â–ª ipAddress is the IP address, entered as a string, of the remote instrument.
See tspnet.connect() (on page 9-399) for additional detail.
To set up communication to a remote ethernet-enabled device that is not TSP enabled:
Send the command:
connectionID = tspnet.connect(ipAddress, portNumber, initString)
Where:
â–ª connectionID is the connection ID used as a handle in all other tspnet function calls.
â–ª ipAddress is the IP address, entered as a string, of the remote device.
â–ª portNumber is the port number of the remote device.
â–ª initString is the initialization string that is to be sent to ipAddress.
See tspnet.connect() (on page 9-399) for additional detail.
To communicate to a remote ethernet device from the 2600B:
1. Connect to the remote device using one of the above procedures. If the 2600B cannot make a
connection to the remote device, it generates a timeout event. Use tspnet.timeout to set the
timeout value. The default timeout value is 20 s.
2. Use tspnet.write() or tspnet.execute() to send strings to a remote device. If you use:
â–ª tspnet.write(): Strings are sent to the device exactly as indicated, and you must supply
any needed termination characters.
â–ª tspnet.execute(): The 2600B appends termination characters to all strings that are sent.
Use tspnet.termination() to specify the termination character.
3. To retrieve responses from the remote instrument, use tspnet.read(). The 2600B suspends
ope2ration until the remote device responds or a timeout event is generated. To check if data is
available from the remote instrument, use tspnet.readavailable().
4. Disconnect from the remote device using the tspnet.disconnect() function. Terminate all
remote connections using tspnet.reset().
Example script
The following example demonstrates how to connect to a remote device that is not TSP
®
enabled,
and send and receive data from this device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals