EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #643 background imageLoading...
Page #643 background image
Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-322 2600BS-901-01 Rev. F/August 2021
Bit
Value and description
B0 to B9
Not used
B10
status.operation.instrument.tsplink.TRIGGER_OVERRUN
status.operation.instrument.tsplink.TRGOVR
Set bit indicates one or more enabled bits for the operation status TSP-Link overrun register
is set.
Bit B10 decimal value: 1,024
Binary value:
0100 0000 0000
B11 to B15
Not used
In addition to the above constants, operationRegister can be set to the numeric equivalent of the
bit to set. For example, to set bit B10, set operationRegister to 1024.
Example
status.operation.instrument.tsplink.enable = 1024
Uses the decimal value to set the trigger overrun bit of the operation status TSP-Link summary enable register.
Also see
Operation Status Registers (on page 12-8)
status.operation.instrument.tsplink.trigger_overrun.* (on page 9-322)
status.operation.instrument.tsplink.trigger_overrun.*
This attribute contains the operation status TSP-Link overrun register set. This command is not available on the
2604B, 2614B, or 2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
14 (All bits set)
Usage
operationRegister =
status.operation.instrument.tsplink.trigger_overrun.condition
operationRegister = status.operation.instrument.tsplink.trigger_overrun.enable
operationRegister = status.operation.instrument.tsplink.trigger_overrun.event
operationRegister = status.operation.instrument.tsplink.trigger_overrun.ntr
operationRegister = status.operation.instrument.tsplink.trigger_overrun.ptr
status.operation.instrument.tsplink.trigger_overrun.enable = operationRegister
status.operation.instrument.tsplink.trigger_overrun.ntr = operationRegister
status.operation.instrument.tsplink.trigger_overrun.ptr = operationRegister
operationRegister
The status of the operation status TSP-link overrun register; a
zero (0) indicates no bits set (also send 0 to clear all bits); other
values indicate various bit settings

Table of Contents

Questions and Answers: