Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-7
TSP commands
The TSP commands available for the instrument are listed in alphabetical order.
beeper.beep()
This function generates an audible tone.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
beeper.beep(duration, frequency)
duration
The amount of time to play the tone (0.001 s to 100 s)
frequency
The frequency of the tone in Hertz (Hz)
Details
You can use the beeper of the 2600B to provide an audible signal at a specified frequency and time
duration. For example, you can use the beeper to signal the end of a lengthy sweep.
The beeper does not sound if it is disabled. It can be disabled or enabled with the beeper enable
command, or through the front panel.
Example
beeper.enable = beeper.ON
beeper.beep(2, 2400)
Enables the beeper and generates a two-second, 2400 Hz tone.
Also see
beeper.enable (on page 9-7)
beeper.enable
This command allows you to turn the beeper on or off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Recall setup
Instrument reset
Saved setup
1 (beeper.ON)
Usage
state = beeper.enable
beeper.enable = state
state
Disable the beeper: beeper.OFF or 0
Enable the beeper: beeper.ON or 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals