Model-specific capabilities
Additional source and measure features:
• Model 2601B, 2602B, and 2604B System SourceMeter
®
instruments:
▪ Source ±dc voltage from 5 µV to 40.4 V
▪ Source ±dc current from 100 pA to 3.03 A
▪ Source ±pulse current up to 10 A
▪ Measure ±pulse current up to 10 A
▪ Measure ±dc voltage from 100 nV to 40.8 V
▪ Measure ±dc current from 100 fA to 3.06 A
• Model 2611B, 2612B, and 2614B System SourceMeter
®
instruments:
▪ Source ±dc voltage from 5 µV to 202 V
▪ Source ±dc current from 2 pA to 1.515 A
▪ Source ±pulse current up to 10 A
▪ Measure ±pulse current up to 10 A
▪ Measure ±dc voltage from 100 nV to 204 V
▪ Measure ±dc current from 100 fA to 1.53 A
• Model 2634B, 2635B, and 2636B System SourceMeter
®
instruments:
▪ Source ±dc voltage from 5 µV to 202 V
▪ Source ±dc current from 1 fA to 1.515 A
▪ Source ±pulse current up to 10 A
▪ Measure ±pulse current up to 10 A
▪ Measure ±dc voltage from 100 nV to 204 V
▪ 2635B and 2636B: Measure ±dc current from 100 aA to 1.53 A
▪ 2634B: Measure ±dc current from 1 fA to 1.53 A
Displaying the serial number
The instrument serial number is on a label on the rear panel of the instrument. You can also access
the serial number from the front panel using the front-panel keys and menus.
To display the serial number on the front panel:
1. If the 2600B is in remote operation, press the EXIT (LOCAL) key once to place the instrument in
local operation.
2. Press the MENU key.
3. Use the navigation wheel to scroll to the SYSTEM-INFO menu item.
4. Press the ENTER key. The SYSTEM INFORMATION menu is displayed.
5. Scroll to the SERIAL# menu item.
6. Press the ENTER key. The 2600B serial number is displayed.