Section 9: TSP command reference  Series 2600B System SourceMeter® instrument Reference Manual 
9-68  2600BS-901-01 Rev. F/August 2021 
 
display.getlastkey() 
This function retrieves the key code for the last pressed key. 
Usage 
keyCode = display.getlastkey() 
 
A returned value that represents the last front-panel key pressed; see Details for 
more information 
 
Details 
A history of the key code for the last pressed front-panel key is maintained by the instrument. When 
the instrument is turned on, or when it is transitioning from local to remote operation, the key code is 
set to 0 (display.KEY_NONE). 
Pressing the EXIT (LOCAL) key normally aborts a script. To use this function with the EXIT (LOCAL) 
key, you must set display.locallockout to display.LOCK. 
 
The table below lists the keyCode value for each front-panel action. 
When using this function, use built-in constants such as display.KEY_RIGHT (rather than the 
numeric value of 103). This allows for better forward compatibility with firmware revisions.