EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #524 background imageLoading...
Page #524 background image
Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-203
Details
This attribute stores the calibration date that is associated with the active calibration set. The
calibration date can be read at any time but can only be assigned a new value when calibration has
been enabled with the smuX.cal.unlock() function. It is typically set to the date when the
instrument was calibrated.
This attribute is stored with the active calibration set. If a different calibration set is restored, this
attribute reflects the date stored with that set.
smuX.cal.date must be set to the date the calibration was done using the UTC time and date. The
date is stored as the number of seconds since UTC 12:00 am Jan 1, 1970.
Due to the internal storage format, smuX.cal.date is accurate to within a few minutes of the
value set.
Example
smua.cal.date = os.time()
Sets calibration date for SMU channel A to the present time set
on the instrument.
Also see
Adjustment (on page 10-15)
os.time() (on page 9-169)
smuX.cal.adjustdate (on page 9-201)
smuX.cal.due (on page 9-203)
smuX.cal.lock() (on page 9-204)
smuX.cal.restore() (on page 9-206)
smuX.cal.save() (on page 9-207)
smuX.cal.state (on page 9-208)
smuX.cal.unlock() (on page 9-209)
smuX.cal.due
This attribute stores the calibration due date for the next calibration.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU cal. restore
SMU nonvolatile memory
0
Usage
calDue = smuX.cal.due
smuX.cal.due = calDue
calDue
Due date of next calibration (0 indicates that no date is set)
X
Source-measure unit (SMU) channel (for example, smua.cal.due applies to
SMU channel A)
Details
This attribute stores the calibration due date associated with the active calibration set. The calibration
due date can be read at any time but can only be assigned a new value when calibration has been
enabled with the smuX.cal.unlock() function. It is typically set to the date when the next
calibration should be performed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals