Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-374 2600BS-901-01 Rev. F/August 2021
trigger.generator[N].assert()
This function generates a trigger event.
Usage
trigger.generator[N].assert()
The generator number (1 or 2)
Details
Use this function to directly trigger events from the command interface or a script. For example, you
can trigger a sweep while the instrument is under script control.
Example
trigger.generator[2].assert()
Generates a trigger event on generator 2.
Also see
trigger.generator[N].EVENT_ID (on page 9-374)
trigger.generator[N].EVENT_ID
This constant identifies the trigger event generated by the trigger event generator.
Usage
eventID = trigger.generator[N].EVENT_ID
The generator number (1 or 2)
Details
This constant is an identification number that identifies events generated by this generator.
To have another trigger object respond to trigger events generated by this generator, set the other
object's stimulus attribute to the value of this constant.
Example
digio.trigger[5].stimulus = trigger.generator[2].EVENT_ID
Uses a trigger event on generator 2 to be the stimulus for digital I/O trigger line 5.