Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-171
printbuffer()
This function prints data from tables or reading buffer subtables.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
printbuffer(startIndex, endIndex, bufferVar)
printbuffer(startIndex, endIndex, bufferVar, bufferVar2)
printbuffer(startIndex, endIndex, bufferVar, ..., bufferVarN)
startIndex
Beginning index of the buffer to print; this must be more than one and less than
endIndex
endIndex
Ending index of the buffer to print; this must be more than startIndex and less
than the index of the last entry in the tables
bufferVar
First table or reading buffer subtable to print
bufferVar2
Second table or reading buffer subtable to print
bufferVarN
The last table or reading buffer subtable to print
...
One or more tables or reading buffer subtables separated with commas
Details
If startIndex ≤ 1, 1 is used as startIndex. If n < endIndex, n is used as endIndex.
When any given reading buffers are used in overlapped commands that have not yet completed (at
least to the specified index), this function outputs data as it becomes available.
When there are outstanding overlapped commands to acquire data, n refers to the index that the last
entry in the table has after all the measurements have completed.
If you pass a reading buffer instead of a reading buffer subtable, the default subtable for that reading
buffer is used.
This command generates a single response message that contains all data. The response message
is stored in the output queue.
The format.data attribute controls the format of the response message.
Example
format.data = format.ASCII
format.asciiprecision = 6
printbuffer(1, rb1.n, rb1)
This assumes that rb1 is a valid reading buffer in the runtime environment. The use of rb1.n
(bufferVar.n) indicates that the instrument should output all readings in the reading buffer. In this example,
rb1.n equals 10.
Example of output data (rb1.readings):
4.07205e-05, 4.10966e-05, 4.06867e-05, 4.08865e-05, 4.08220e-05, 4.08988e-05,
4.08250e-05, 4.09741e-05, 4.07174e-05, 4.07881e-05

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals