Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-335
status.questionable.calibration.*
This attribute contains the questionable status calibration summary register set.
2601B, 2611B, 2635B: 2 (All bits set)
Models 2602B, 2604B, 2612B, 2614B,
2634B, 2636B: 6 (All bits set)
Usage
questionableRegister = status.questionable.calibration.condition
questionableRegister = status.questionable.calibration.enable
questionableRegister = status.questionable.calibration.event
questionableRegister = status.questionable.calibration.ntr
questionableRegister = status.questionable.calibration.ptr
status.questionable.calibration.enable = questionableRegister
status.questionable.calibration.ntr = questionableRegister
status.questionable.calibration.ptr = questionableRegister
The status of the questionable status calibration summary
register; a zero (0) indicates no bits set (also send 0 to clear all
bits); other values indicate various bit settings
Details
These attributes are used to read or write to the questionable status calibration summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register set
contents (on page 12-1) and Enable and transition registers (on page 12-19). The individual bits of
this register are defined in the following table.
status.questionable.calibration.SMUA
Set bit indicates that the SMU A calibration constants stored in nonvolatile memory were
corrupted and could not be loaded when the instrument powered up.
Bit B1 decimal value: 2
Binary value: 0000 0010
status.questionable.calibration.SMUB
This bit is only available on 2602B, 2604B, 2612B, 2614B, 2634B, 2636B. Set bit indicates that
the SMU B calibration constants stored in nonvolatile memory were corrupted and could not be
loaded when the instrument powered up.
Bit B2 decimal value: 4
Binary value: 0000 0100