EasyManua.ls Logo

Tektronix KEITHLEY 2600B Series

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Series 2600B System SourceMeter® instrument Reference Manual Section 10: Calibration
2600BS-901-01 Rev. F/August 2021 10-31
Step 4. Contact check calibration adjustment
Skip this step for the 2604B, 2614B, and 2634B. These models do not perform contact check
measurements.
A. As illustrated in the following figure:
Short the 2600B SENSE LO and LO terminals together.
Short the SENSE HI and HI terminals together.
Figure 135: Connections for contact check 0 ohm calibration
B. Allow the readings to settle, then get the 2600B readings:
r0_hi, r0_lo = smua.contact.r()
C. Characterize both 50 Ω resistors using the resistance function of the digital multimeter.
D. As illustrated in the following figure:
Characterize both 50 Ω resistors using the resistance function of the digital multimeter.
Connect a 50 Ω resistor between the SENSE LO and LO terminals.
Connect the second 50 Ω resistor between the SENSE HI and HI terminals.
Figure 136: Connections for contact check 50 ohm calibration
E. Allow the readings to settle, then get the 2600B readings:
r50_hi, r50_lo = smua.contact.r()

Table of Contents

Related product manuals