Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-88 2600BS-901-01 Rev. F/August 2021
errorqueue.clear()
This function clears all entries out of the error queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
errorqueue.clear()
Details
See Error queue (on page 12-3) for additional information about the error queue.
Also see
errorqueue.count (on page 9-88)
errorqueue.next() (on page 9-89)
errorqueue.count
This attribute gets the number of entries in the error queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Power cycle
Clearing error queue
Reading error messages
Not applicable
Not applicable
Usage
count = errorqueue.count
count
The number of entries in the error queue
Example
count = errorqueue.count
print(count)
Returns the number of entries in the error queue.
The output below indicates that there are four entries in the error queue:
4.00000e+00
Also see
Error queue (on page 12-3)
errorqueue.clear() (on page 9-88)
errorqueue.next() (on page 9-89)