Section 6: Troubleshooting guide Series 2600B System SourceMeter® instrument Reference Manual
6-6 2600BS-901-01 Rev. F/August 2021
Error summary
Error number
Error level
Error message
5048
RECOVERABLE
Contact check not valid with HIGH-Z OUTPUT off
5049
RECOVERABLE
Contact check not valid while an active current source
5050
RECOVERABLE
I limit too low for contact check
5051
FATAL
Model number/SMU hardware mismatch. Disconnect DUT and cycle power
5052
RECOVERABLE
Interlock engaged; system stabilizing
5053
RECOVERABLE
Unstable output detected - Measurements may not be valid
5055
RECOVERABLE
Cannot change adjustment date - change cal constants before operation
5059
RECOVERABLE
trigger.source.action enabled without configuration
5060
RECOVERABLE
trigger.measure.action enabled without configuration
5061
RECOVERABLE
Operation not permitted while OUTPUT is off
5062
SERIOUS
SMU overload. Automatic OUTPUT off.
5063
RECOVERABLE
Cannot perform requested operation while measure autozero is on
5064
RECOVERABLE
Cannot use reading buffer that collects source values
5065
RECOVERABLE
I range too low for contact check
5066
RECOVERABLE
source.offlimiti too low for contact check
5069
SERIOUS
Autorange locked for HighC mode
LAN troubleshooting suggestions
If you are unable to connect to the web interface of the instrument, check the following items:
• Verify that the network cable is in the LAN port on the rear panel of the instrument, not one of the
TSP-Link
®
ports (see the description in Rear panel).
• Verify that the network cable is in the correct port on the computer. The LAN port of a laptop may
be disabled when the laptop is in a docking station.
• Verify that the configuration information for the correct ethernet card was used during the setup
procedure.
• Verify that the network card in the computer is enabled.
• Verify that the IP address of the instrument is compatible with the IP address on the computer.
• Verify that the subnet mask address of the instrument is the same as the subnet mask address of
the computer.
• Turn the instrument power off, and then on. Wait at least 60 seconds for the network
configuration to be completed. Verify that an IP address has been assigned to the instrument:
1. Press the MENU key to display the MAIN MENU.
2. Use the navigation wheel to select LAN. The LAN CONFIG menu is displayed.
3. Select STATUS.
4. Select IP-ADDRESS.
• Restart your computer.
• For more detail on LAN settings, see Connecting to the LAN (on page 8-42).
If the above actions do not correct the problem, contact your system administrator.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals