Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-346 2600BS-901-01 Rev. F/August 2021
Example
requestEventRegister = status.request_event
print(requestEventRegister)
Reads the status request event register.
Sample output:
1.29000e+02
Converting this output (129) to its binary equivalent yields 1000 0001.
Therefore, this output indicates that the set bits of the status request event register are presently B0 (MSB) and
B7 (OSB).
Also see
status.condition (on page 9-280)
status.system.* (on page 9-349)
Status byte and service request (SRQ) (on page 12-15)
status.reset()
This function resets all bits in the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
status.reset()
Details
This function clears all status data structure registers (enable, event, NTR, and PTR) to their default
values. For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status
register set contents (on page 12-1) and Enable and transition registers (on page 12-19).
Example
status.reset()
Resets the instrument status model.
Also see
Status model (on page 12-1)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals