Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-346 2600BS-901-01 Rev. F/August 2021
requestEventRegister = status.request_event
print(requestEventRegister)
Reads the status request event register.
Sample output:
1.29000e+02
Converting this output (129) to its binary equivalent yields 1000 0001.
Therefore, this output indicates that the set bits of the status request event register are presently B0 (MSB) and
B7 (OSB).
Also see
status.condition (on page 9-280)
status.system.* (on page 9-349)
Status byte and service request (SRQ) (on page 12-15)
status.reset()
This function resets all bits in the status model.
Usage
status.reset()
Details
This function clears all status data structure registers (enable, event, NTR, and PTR) to their default
values. For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status
register set contents (on page 12-1) and Enable and transition registers (on page 12-19).
Example
Resets the instrument status model.
Also see
Status model (on page 12-1)