Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-309
status.operation.instrument.lan.trigger_overrun.*
This attribute contains the operation status LAN trigger overrun register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
510 (All bits set)
Usage
operationRegister = status.operation.instrument.lan.trigger_overrun.condition
operationRegister = status.operation.instrument.lan.trigger_overrun.enable
operationRegister = status.operation.instrument.lan.trigger_overrun.event
operationRegister = status.operation.instrument.lan.trigger_overrun.ntr
operationRegister = status.operation.instrument.lan.trigger_overrun.ptr
status.operation.instrument.lan.trigger_overrun.enable = operationRegister
status.operation.instrument.lan.trigger_overrun.ntr = operationRegister
status.operation.instrument.lan.trigger_overrun.ptr = operationRegister
operationRegister
The status of the operation status LAN trigger overrun register; a zero (0) indicates
no bits set (also send 0 to clear all bits); other values indicate various bit settings
Details
These attributes are used to read or write to the operation status LAN trigger overrun registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15. For example, if a value of 2.58000e+02 (which is 258) is read as the value of the
condition register, the binary equivalent is 0000 0001 0000 0010. This value indicates that bit B1 and
bit B8 are set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
1
0
0
0
0
0
0
1
0
* Least significant bit
** Most significant bit
A set bit indicates that the specified LAN trigger generated an action overrun when triggered to
generate a trigger packet.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register set
contents (on page 12-1) and Enable and transition registers (on page 12-19). The individual bits of
this register are defined in the following table.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals