EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #639 background image
Section 9: TSP command reference Series 2600B System SourceMeter® instrument Reference Manual
9-318 2600BS-901-01 Rev. F/August 2021
status.operation.instrument.trigger_timer.*
This attribute contains the operation status trigger timer summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
1024 (All bits set)
Usage
operationRegister = status.operation.instrument.trigger_timer.condition
operationRegister = status.operation.instrument.trigger_timer.enable
operationRegister = status.operation.instrument.trigger_timer.event
operationRegister = status.operation.instrument.trigger_timer.ntr
operationRegister = status.operation.instrument.trigger_timer.ptr
status.operation.instrument.trigger_timer.enable = operationRegister
status.operation.instrument.trigger_timer.ntr = operationRegister
status.operation.instrument.trigger_timer.ptr = operationRegister
operationRegister
The status of the operation status trigger timer summary register; a zero (0) indicates
no bits set (also send 0 to clear all bits); the only valid value other than 0 is 1024
Details
These attributes are used to read or write to the operation status trigger timer summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register set
contents (on page 12-1) and Enable and transition registers (on page 12-19). The individual bits of
this register are defined in the following table.
Bit
Value and description
B0 to B9
Not used
B10
status.operation.instrument.trigger_timer.TRIGGER_OVERRUN
status.operation.instrument.trigger_timer.TRGOVR
Set bit indicates one or more enabled bits for the operation status trigger timer overrun register is
set.
Bit B10 decimal value: 1,024
Binary value:
0100 0000 0000
B11 to B15
Not used
In addition to the above constants, operationRegister can be set to the numeric equivalent of the
bit to set. For example, to set bit B10, set operationRegister to 1024.
Example
status.operation.instrument.trigger_timer.enable = 1024
Uses the decimal value to set the TRGOVR bit of the operation status trigger timer summary enable register.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals