EasyManuals Logo
Home>Tektronix>Measuring Instruments>KEITHLEY 2600B Series

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #821 background imageLoading...
Page #821 background image
Section 14: Model 2400 emulation Series 2600B System SourceMeter® instrument Reference Manual
14-14 2600BS-901-01 Rev. F/August 2021
Subsystem
Command
Supported
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut
SOURce|DELay|SENSe|NONE
Partially
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut?
Yes
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:OLINe 1|2|3|4
Partially
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:OLINe?
Yes
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:OUTPut
SOURce|DELay|SENSe|NONE
Yes
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:OUTPut?
Yes
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection SOURce|ACCeptor
Yes
:TRIGger
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection?
Yes
:TRIGger
:TRIGger:SEQuence2:SOURce <name>
No
:TRIGger
:TRIGger:SEQuence2:SOURce?
No
:TRIGger
:TRIGger:SEQuence2:TOUT <NRf>|DEFault|MINimum|MAXimum
No
:TRIGger
:TRIGger:SEQuence2:TOUT? [DEFault|MINimum|MAXimum]
No
Model 2400 SCPI command compatibility
The following tables lists all the Model 2400 SCPI commands that are not fully supported by the
Model 2400 personality script (Persona2400) and specific compatibility details for each command.
:CALCulate subsystem
:CALCulate2:NULL:ACQuire
Each time the personality script is run, the null offset value is reset to zero (0). Until a measurement is made, this
command uses a zero (0) for the null offset value. After a measurement is made, this command uses latest
measurement to set the null offset value (the same method used as a Model 2400).
:DISPlay subsystem
:DISPlay:DIGits 4|5|6|7|DEFault|MINimum|MAXimum
When 3.5 digits or MINimum is requested, the instrument sets the resolution to 4.5 digits.
:DISPlay:DIGits? [DEFault|MINimum|MAXimum]
When queried for the MINimum, the instrument returns 5.
:DISPlay:ENABle <Bool>
The Series 2600B performance is not degraded by display operation. The instrument accepts the command but no
action is performed.
:DISPlay:WINDow2:ATTRibutes?
The instrument always returns 32 zeros.
:DISPlay:WINDow2:TEXT:STATe <Bool>
Changing this setting also changes the :DISPlay[:WINDow[1]]:TEXT:STATe setting to the same value.
DISPlay[:WINDow[1]]:ATTRibutes?
The instrument always returns 20 zeros.
DISPlay[:WINDow[1]]:TEXT:STATe <Bool>
Changing this setting also changes the :DISPlay:WINDow2:TEXT:STATe setting to the same value.

Table of Contents

Questions and Answers: