Series 2600B System SourceMeter® instrument Reference Manual Section 5: Theory of operation
2600BS-901-01 Rev. F/August 2021 5-5
Figure 61: 2601B, 2602B, or 2604B continuous power operating boundaries
The general operating boundaries for 2611B, 2612B, 2614B, 2634B, 2635B, or 2636B continuous
power output are shown in the following figure (for derating factors, see General power equation (on
page 5-3)). In this figure, the current (100 mA and 1.5 A) and voltage (20 V and 200 V) magnitudes
are nominal values. Also note that the boundaries are not drawn to scale.
Figure 62: 2611B, 2612B, 2614B, 2634B, 2635B, or 2636B continuous power operating
boundaries