EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #684 background imageLoading...
Page #684 background image
Series 2600B System SourceMeter® instrument Reference Manual Section 9: TSP command reference
2600BS-901-01 Rev. F/August 2021 9-363
Details
Data for current measurements, voltage source values, and timestamps are stored in
smuX.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters
set to the default values.
Performs a linear voltage sweep with current measured at every step (point):
1. Sets the SMU to output startv amperes, allows the source to settle for stime seconds, and
then makes a current measurement.
2. Sets the SMU to output the next amperes step, allows the source to settle for stime seconds,
and then makes a voltage measurement.
3. Repeats the above sequence until the voltage is measured on the stopv amperes step.
The linear step size is automatically calculated as follows:
step = (stopv – startv) / (points – 1)
Figure 127: SweepVLinMeasureI()
Example
SweepVLinMeasureI(smua, -1, 1, 1e-3, 1000)
This function performs a 1000-point linear voltage sweep starting at -1 V and stopping at +1 V. Current is
measured at every step (point) in the sweep after a 1 ms source settling period.
Also see
KISweep factory script (on page 7-22)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals