EasyManuals Logo

Tektronix KEITHLEY 2600B Series User Manual

Tektronix KEITHLEY 2600B Series
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #746 background imageLoading...
Page #746 background image
Section 10: Calibration Series 2600B System SourceMeter® instrument Reference Manual
10-10 2600BS-901-01 Rev. F/August 2021
6. Verify measure current accuracy for each of the currents listed using the values listed in the
following table for your model number. For each measurement:
â–ª Select the correct source range.
â–ª Set the 2600B output current such that the digital multimeter reading is the value indicated in
the source current column of the table below. It may not be possible to set the current source
to get exactly the required reading on the digital multimeter. Use the closest possible setting
and modify the reading limits accordingly.
â–ª If necessary, press the TRIG key to display readings.
â–ª Verify that the 2600B current reading is within the limits given in the table below.
7. Repeat the procedure for negative calibrator currents with the same magnitudes as those listed.
2601B, 2602B, and 2604B current measurement accuracy limits
Source and measure range
1
Source current
2
Current reading limits (1 year,
18° C to 28° C)
100 nA
90.000 nA
89.855 nA to 90.145 nA
1 µA
0.9000 µA
0.89928 µA to 0.90073 µA
10 µA
9.0000 µA
8.9963 µA to 9.0038 µA
100 µA
90.000 µA
89.957 µA to 90.043 µA
1 mA
0.9000 mA
0.89962 mA to 0.90038 mA
10 mA
9.0000 mA
8.9957 mA to 9.0043 mA
100 mA
90.000 mA
89.962 mA to 90.038 mA
1 A
0.90000 A
0.89823 A to 0.90177 A
3 A
2.4000 A
2.3953 A to 2.4047 A
1. Measure range coupled to source range when simultaneously sourcing and measuring current.
2. As measured by precision digital multimeter. Use closest possible value and modify reading limits
accordingly if necessary. See Measurement limit calculations (on page 10-4).
2611B, 2612B, and 2614B current measurement accuracy limits
Source and measure range
1
Source current
2
Current reading limits (1 year,
18° C to 28° C)
100 nA
90.000 nA
89.846 nA to 90.154 nA
1 µA
0.9000 µA
0.89928 µA to 0.90073 µA
10 µA
9.0000 µA
8.9963 µA to 9.0038 µA
100 µA
90.000 µA
89.957 µA to 90.043 µA
1 mA
0.9000 mA
0.89962 mA to 0.90038 mA
10 mA
9.0000 mA
8.9957 mA to 9.0043 mA
100 mA
90.000 mA
89.962 mA to 90.038 mA
1 A
0.90000 A
0.89823 A to 0.90177 A
1.5 A
1.3500 A
1.34583 A to 1.35418 A
1. Measure range coupled to source range when simultaneously sourcing and measuring current.
2. As measured by precision digital multimeter. Use closest possible value and modify reading limits
accordingly if necessary. See Measurement limit calculations (on page 10-4).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals