Series 2600B System SourceMeter® instrument Reference Manual Section 10: Calibration
2600BS-901-01 Rev. F/August 2021 10-19
2601B, 2602B, or 2604B calibration steps
Function
1
Calibration steps
2
Calibration points
4
Sense mode
5
Voltage Source
and Measure
100 mV
100 mV
1 V
1 V
6 V
40 V
±1e-30, ±90 mV
±1e-30, ±90 mV
±1e-30, ±0.9 V
±1e-30, ±0.9 V
±1e-30, ±5.4 V
±1e-30, ±36 V
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
Current Source and Measure
100 nA
1 μA
10 μA
100 μA
1 mA
1 mA
10 mA
100 mA
1 A
3 A
10 A
3
±1e-30, ±90 nA
±1e-30, ±0.9 μA
±1e-30, ±9 μA
±1e-30, ±90 μA
±1e-30, ±0.9 mA
±1e-30, ±0.9 mA
±1e-30, ±9 mA
±1e-30, ±90 mA
±1e-30, ±0.9 A
±1e-30, ±2.4 A
±1e-30, ±2.4 A
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
1. Calibrate only the source for the CALA sense steps.
2. Steps must be performed in the order shown.
3. 10 A range for changing calibration of range only and is not available for normal use.
4. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-30. Calibration polarities must also be
set as shown in the procedures.
5. Output must be off before changing to the CALA sense mode.
2611B, 2612B, or 2614 calibration steps
Function
1
Calibration steps
2
Calibration points
3
Sense mode
4
Voltage Source
and Measure
200 mV
200 mV
2 V
2 V
20 V
200 V
±1e-30, ±180 mV
±1e-30, ±180 mV
±1e-30, ±1.8 V
±1e-30, ±1.8 V
±1e-30, ±18 V
±1e-30, ±180 V
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
Current Source and Measure
100 nA
1 μA
10 μA
100 μA
1 mA
1 mA
10 mA
100 mA
1 A
1.5 A
10 A
±1e-30, ±90 nA
±1e-30, ±0.9 μA
±1e-30, ±9 μA
±1e-30, ±90 μA
±1e-30, ±0.9 mA
±1e-30, ±0.9 mA
±1e-30, ±9 mA
±1e-30, ±90 mA
±1e-30, ±0.9 A
±1e-30, ±1.35 A
±1e-30, ±2.4 A
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
1. Calibrate only the source for the CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-30. Calibration polarities must also be
set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals