Typical values for the 100 mA range:
smua.source.calibratei(-100e-3, -1e-30, -1e-6, -90e-3, -89.2e-3)
22. If this step is not on the CALA sense mode, send the measure calibration command using the
multimeter and 2600B readings, and range setting for the parameters:
smua.measure.calibratei(-range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg)
Where:
Typical 100 mA range values:
smua.measure.calibratei(-100e-3, -1e-5, -1e-6, -91e-3, -89.2e-3)
D. Before continuing, be sure to complete steps 1 through 22 for the 100 nA to 1 A ranges before
continuing with 3 A and 10 A range calibration (2601B, 2602B, 2604B) or 1.5 A and 10 A range
calibration (2611B, 2612B, 2614B, 2634B, 2635B, 2636B).
E. Change connections as shown in the following figure.
Figure 134: Connections for 1.5 A and 3 A current ranges
F. Select the DMM dc volts function.
G. Repeat the 22 steps of C for the 3 A and 10 A ranges (2601B, 2602B, 2604B) or 1.5 A and 10 A
ranges (2611B, 2612B, 2614B, 2634B, 2635B, 2636B). Compute the current reading from the
DMM voltage reading and characterized 0.5 Ω resistance value: I = V/R.
H. Select automatic polarity mode:
smua.cal.polarity = smua.CAL_AUTO