Specifications
Hardware Description
2-49
Environmental The following table gives information on the environmental operating requirements for the
current system components. Data for both the Philips system as a whole and for individual
components are provided.
Table 2-5. Environmental Requirements for Information Center System Components
Product
Part #
Temperature
Relative Humidity
(Non-condensing)
Philips System 59 - 86
o
F
15 - 30
o
C
20 - 80% @ 30
o
C (86
o
F)
Philips Components
Server (M3154) M3168A 5 - 35
o
C
(41 - 95
o
F)
5 - 95%
PC M3167B 10 - 35
o
C
(50 - 95
o
F)
10 - 90%
Medium CRT Display M3161 32 - 104
o
F
0 - 40
o
C
8 - 85%
Medium CRT Display - Europe M3161 32 - 104
o
F
0 - 40
o
C
8 - 85%
Large CRT Display M3162 32 - 104
o
F
0 - 40
o
C
10 - 80%
Large Flat Panel Display 862103
Medium Touch Flat Panel Display 862428 50 - 104
o
F
10 - 40
o
C
0 - 80%
Video Splitter 2-way
6-way
862096
862097
41 - 104
o
F
5 - 40
o
C
15 - 80%
LaserJet Printer 2300L 59 - 89
o
F
15 - 32.5
o
C
10 - 80%
Printer Hub AT-FS708 32 - 104
o
F
0 - 40
o
C
5 - 95%
Philips 2-Channel Recorder/Rack M1276A #201/
M1116B
41 - 104
o
F
5 - 40
o
C
15 - 80%
2-Channel Recorder/Rack Power Supply M3180-60040 41 - 104
o
F
5 - 40
o
C
15 - 80%
4-Channel Recorder/Power Supply M3160A 41 - 104
o
F
5 - 40
o
C
20 - 80%
UPS (650 VA,120V) 862099 41 - 104
o
F
5 - 40
o
C
15 - 80%
UPS (650 VA, 230V) 0950-3406
UPS for Database Server
(1000 VA, 120V)
M3882-60001 32 - 104
o
F
0 - 40
o
C
0 - 95%
UPS for Database Server (Japan)
(1000 VA, 100V)
M3884-60001 32 - 104
o
F
0 - 40
o
C
0 - 95%